Clearview Fume Hood (BC 6003)

24 inch wide Ductless Fume Hood with recirculation filtration providing operator and environmental protection from chemical fumes, vapours and particulates (when HEPA filtration selected) during laboratory procedures.

This hood has unobstructed view into the hood interior including through the back panel and is fully portable, lighting from a white LED light aids the internal brightness and comfort for the operator.Hood has double folding acrylic front windows to permit full width access for placing apparatus within the hood. Services ports in both side windows provide access for cable and hose connections. The included PVC base spillage tray has excellent chemical resistance and contains any spills.

Hood uses F3 sized filter and has teardrop front working access as standard (alternative trapezium or other shaped front access on request) with double hinged and lockable front door flaps for easy access to the interior. Each side panel has a services port. Hoods can be readily installed onto existing benching or supplied with a static/mobile bench having optional cupboard doors and services kit. This bench can be supplied with adjustable leveling feet or large lockable castors.

Key Features:

  • This fume hood offers excellent unhindered 360° all-round vision.
  • Safe-change pre-filter can be bagged under negative pressure.
  • Fully portable.
  • Keypad control panel with touch button controls and service due indicator.
  • Audible low airflow alarm.
  • The work surface is black which highlights spillages.
  • Quiet operation, <50d(B)A.
  • Lockable front access doors.
  • Average face velocity of 98 fpm.


Standards
(Chemcap conform to the requirements of all major Standards for containment and filtration):

  • US Standard ASHRAE 110: 1995 12-03 for containment.
  • British Standard BS 7989:2001 for filter efficiency and retention capacity.
  • French Standard AFNOR NFX 15-211 – Class 2 for filter retention capacity.
  • Canadian Standard CSA Z316.5 for containment.


Optional Features:

  • Electronic Carbon Filter Saturation Alarm.
  • Rear Baffle.
  • Top Hat Exhaust Spigot for connection to house ducting.
  • Upgrade base tray to Stainless Steel.
  • Stainless Steel Heat/flame Protection Shield for filters.
  • Swan neck water tap and hose and Oval drip cup with drain hose.
  • Services Kit – gas tap, swan neck water tap, oval drip cup all with drain hose and double electrical socket.
  • Moblie or Static Bench as seen below.


Bench Options:
Either the static or moblie style of bench may be supplied with 5″ castors or levelling/adjustable feet.

Chemcap-Clearview-mobile-bench Chemcap-Clearview-ductless-fume-hoods-EU-5
SEND ENQUIRY ABOUT THIS PRODUCT

Clearview Fume Hood (BC 6003)

Construction and Materials:
The mild steel fan housing is epoxy coated, acid resistant paint, color – white.
Transparent panels, including hinged doors, are manufactured from thick clear cast acrylic which is non-flammable and easy to clean.
The base tray is manufactured from PVC, color – black.

 Hood Size & Model Number:

Model Description Dimensions
(External WxDxH)
Dimensions
(Internal WxDxH)
BC 6003 BC 6003 | Chemcap Clearview Ductless Fume Cabinet supplied complete with LED lighting and filtration as shown on flitration tab. Nominal 23.5″ x 22.5″ x 45″
(Design 600 x 570 x 1145mm)
Nominal 22.5″ x 20.5″ x 32″
(Design 570 x 520 x 810mm)


Bench Size & Part Number:

Part No. Description Dimensions mm
(External WxDxH)
MT 6000 Bench robust epoxy coated mild steel construction. Nominal 24″ x 23″ x 45″
(Design 600 x 570 x 1145mm)
MTC 6000 Mobile bench robust epoxy coated mild steel construction, doors with magnetic catches and lockable 130mm castors. Nominal 24″ x 23″ x 45″
(Design 600 x 570 x 1145mm)

 

Hoods will be supplied for power supply: 110V, AC, 60Hz, 8Amp, 1Ø unless 230V, AC, 50Hz, 5Amp, 1Ø is specifically requested.

SEND ENQUIRY ABOUT THIS PRODUCT

Clearview Fume Hood (BC 6003)

Filtration:
Chemcap clearview hoods utilize carbon filtration with or without HEPA particulate filtration:

Chemcap-Filtration-Diagrams

Choose your optimised Chemcap filtration with or without particulate filtration according to the uses for your hood:

chemcap-filtration-table


Free Risk Assessment Service:
What are your risks? Please contact us for a free assessment and our chemist will undertake a risk analysis and suggest the best filter for your purposes. Please click on the download button to the right.

SEND ENQUIRY ABOUT THIS PRODUCT